Force NC

PTCA Non-Compliant Balloon Catheter

Precise Powerful Post Dilatation

About FORCE NC

FORCE NC, the high pressure PTCA catheter that ensures pressure when pressure is needed without compromising on flexibility, profile and catheter performance. Now get the catheter with superb performances that enables precise and safe post dilatation.

THE LOW PROFILE, FLEXIBLE HIGH PRESSURE PTCA

High rated burst pressure

  • Rated burst pressure 21 bar (2,0-3,0 mm), nominal pressure 12 bar
  • Providing an extensive and safe working range

Conformability

  • No longitudinal growth for controlled expansion, even at high pressures
  • No balloon deformations up to rated burst pressure for precise stent apposition
  • No dog boning between nominal and rated burst pressure

Flexibility and low profile

  • Optimized materials and design for flexibility and low profiles
  • Ease of use equal to high performance workhorse catheters

Real non compliantReal non compliant

  • Real non compliant: ~1% growth per atmosphere between nominal and rated burst pressure
  • Controlled equal growth during inflation up to rated burst pressure
  • Precise compliance ensuring exact and protected sizing

Hydrophilic coating

  • Fully coated, tip to distal shaft including folded balloon
  • Reduced likehood of catheter friction enchances ease of passage
  • Remains durable during multiple inflations

Technical Data

Nominal Pressure: 12 bar
Rated Burst Pressure: Ø 2.00 – 3.00 = 21 bar

Ø 3.25 – 4.50 = 19 bar

Balloon material: Non-compliant polyamide for

controlled dimensions

Folding: 3-folds
Guiding catheter compatibility: 5F (min ID 0.056”/1.42 mm)
Guide wire compatibility: max 0.014” (0.36 mm)
Catheter type: Rapid Exchange
Proximal shaft: 1.9F (0.64 mm)
Distal shaft: 2.7F (0.90 mm)
Tip entry profile: 0.016” (0.40 mm)
Usable Catheter Length: 154 cm
Tapered tip: 3.5 mm
Shaft coating: Hydrophilic coating
Shelf life: 3 Years from sterilization date